GE MEDICAL SYSTEMS
REV 5
ILLUSTRATION 10-11
COLLIMATOR CABLE ROUTE ON CATHODE CABLE
ILLUSTRATION 10-12
COLUMN TERMINAL STRIP AMX3 A1 TS1
3. There is a large loop of collimator cable immediately after strap point C." See
Illustration 10-11.
CLAMP POINT D
STRAP POINT N
4. The collimator cable goes from strap point C" directly into the vertical column
assembly terminal box where it is connected to terminal strip AMX3 A1 TS1. See
Illustration 10-12 and Table 10-1.
TERMINAL #1
VERTICAL COLUMN
BRAKE CABLE
WIRE CABLES
PER CHART
COLUMN
TERMINAL BOX
TERMINAL #16
GND STUD
HORIZONTAL ARM BRAKE
AMX3 A2 J1
X-RAY TUBE
AMX3 A3 TS1
10-9
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
STRAP POINT C
CLAMP POINT A
COLLIMATOR
AMX3 A4
X-RAY CONTROL
AMX1 A2 TS3
AMX-4+ SERVICE