GE MEDICAL SYSTEMS
REV 5
ILLUSTRATION 1-3
PORT AND COLLIMATOR INTERFACE PLATE INSTALLATION
MOBILE SYSTEMS USE EITHER
FOUR OR EIGHT 1/4-20 X 0.625 INCH
PORT PLATE SCREWS
TUBE
HOUSING
PORT PLATE
TUBE
HOUSING
(WITH COLLIMATOR REMOVED)
INTERFACE
PLATE
6. Remove collimator from interface plate.
7. Carefully place collimator and cables on horizontal arm.
8. Check that each X-ray port plate screw is securely fastened in place.
9. If any port plate screw is loose, unscrew it until it touches the interface plate.
1-3
(MODEL 2169360, 2236420 & 2275938 SERIES)
10-32 X 0.50 INCH
INTERFACE PLATE
SECTION A"
AMX-4+ SERVICE
DIRECTION 2173225-100
SCREWS