Replacing The Compact Optics; Camera Optics - Siemens SIREMOBIL Iso-C Service Instructions Manual

Hide thumbs Also See for SIREMOBIL Iso-C:
Table of Contents

Advertisement

Replacing boards / replacing components
Fig. 6

Replacing the Compact optics

WARNING
Removal
WARNING
CAUTION

Camera optics

Move the I.I. into the service position. Refer to chapter 1, I.I. service position.
Rotate the camera into the 0° position. Refer to Fig. 6.
Siemens AG
Medical Solutions
0° Position
Electrical voltage!
See Chapter 1, Safety Information.
Switch the SIREMOBIL Compact off and wait approximately
3 minutes for the voltage in the I.I. mini voltage supply to drop.
Risk of injuries on mechanical parts!
If not observed, slight to medium injuries, especially to the hands,
can occur.
Observe the notes concerning the C-arm carbon fiber structure.
Risk of injury!
Noncompliance can lead to light to medium injuries and/or
property damage.
The C-arm covers contain counterbalance weights and weigh
approx. 6 kg.
Hold the covers firmly when installing or removing them.
Before the C-arm cover will be loosened, the C-arm must be
moved to its middle orbital position.
If the C-arm covers are loosened, the C-arm may not be moved to
its orbital end positions (I.I. or SIREPHOS at the orbital end posi-
tion).
SPR2-230.061.01
Rev. 13
12.04
Page 17 of 48
SIREMOBIL Iso-C
CS PS 24
6 - 17
6
6
System Manual

Advertisement

Table of Contents
loading

Table of Contents