Replacing The Compact Optics; Removing The Compact Optics - Siemens SIREMOBIL Compact L Service Instructions Manual

Table of Contents

Advertisement

6 - 10
1
2
Fig. 5
1
3
2
Fig. 7

Replacing the Compact optics

WARNING

Removing the Compact optics

Turn the C-arm so that the I.I. is at the top.
Loosen and remove the I.I. cover.
Rotate the camera in the "0" position (6/Fig. 6).
SIREMOBIL Compact
System Manual
Replacing boards / replacing components
4
3
5
3
4
5
3
2
4
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the mini voltage supply, the system must be
switched off for at least 3 minutes until the high voltage in the
system and in the I.I. mini voltage supply dissipates.
SPR2-130.061.02
Rev. 11
6
5
5
3
Fig. 6
Fig. 8
.
Page 10 of 28
09.04
CS PS 24
2
5
4
3
1
1
2
3
Siemens AG
Medical Solutions
6
6

Hide quick links: