6 - 26
MEMOSKOP C-E 100
NOTE
NOTE
Replace the MEMOSKOP C-E 100.
The attachment screws are accessible on the underside of the storage device.
Check the software status of the MEMOSKOP. Download the system software to the
MEMOSKOP if necessary.
Use the MEMOSKOP Service Software to import the language file.
Set the MEMOSKOP C-E 100 to the correct video frequency and configure it for the
number of existing monitors.
Refer to chapter 5 "Configuring the video frequency " and "MEMOSKOP - Configuration
for the number of existing monitors ."
Configure the following parameters in Technical Setup for MEMOSKOP:
Number of Monitors
Hardcopy
Edge 1 coef set
Edge 2 coef set
Motion detection setup
Date / Time format
Keyboard
Language
Display of realtime clock
Display of rotated images
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
Check the software status of the newly delivered MEMOSKOP.
If the software status of the MEMOSKOP differs from that of the
system, the system software status must be downloaded to the
MEMOSKOP.
The MEMOSKOP software status can be read out in the 2nd line
of the User Setup.
The first number describes the MEMOSKOP type.
(2=CE; 3= CE-100/ C/ C & MOD; 4=C-SUB/ C-SUB & MOD).
The image storage MEMOSKOP C-E 100 is a new image storage
that works only with the correct number of images. For this rea-
son, if the image storage was set to its default values
(Menu "Diagnostics", "Load Factory Defaults"), the parameter
"Maximum image numbers" must be set to 100 images.
Read and record the parameters for MEMOSKOP C-E 100 if you
can select User setup. After replacing the MEMOSKOP, configure
the MEMOSKOP C-E 100 according to the parameters noted.
Configure 1 for SIREMOBIL with 1 monitor.
Configure 2 for SIREMOBIL with 2 monitors.
Check whether 5000 is programmed in "4 Max Hardcopy
time". If not, program this value.
Check whether 3 is programmed, If not, program this value.
Check whether 8 is programmed, If not, program this value.
Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not, pro-
gram these values.
Set the country-specific format.
Set to the keyboard layout being used.
Set for language of the country.
Configure Yes. Select No only upon customer request.
Configure Yes.
SPR2-230.061.01
Rev. 13
Page 26 of 48
12.04
CS PS 24