ARTD-002.731.17. The protective conductor resistance must not exceed 0.2 Ohm. Validity of this document This document is applicable if a key switch is to be retrofitted to the SIREMOBIL Compact, SIREMOBIL Compact L, SIREMOBIL Iso-C, POWERMOBIL, and ARCOSKOP systems. Documents required •...
Drill a 3.5 mm hole at the marked position (Fig. 1). • Increase the hole diameter to 22 mm with the step drill (Fig. 1). • Remove the remaining drilling chips in the monitor trolley with a vacuum cleaner. Siemens AG SPR2-130.812.06 Page 3 of 6 SIREMOBIL Medical Solutions Rev.
Page 4
• Tighten the screw (1/Fig. 3) clockwise, until the key switch is fixed firmly in the hole. • Do not yet install the switch top (4/Fig. 4). SIREMOBIL SPR2-130.812.06 Page 4 of 6 Siemens AG System Manual Rev.
Page 5
• Unplug connector D50.X5 on board D50. • With the SIREMOBIL Iso-C, POWERMOBIL, and ARCOSKOP systems, the cable is ad- ditionally equipped with the connector X5B. Remove it from the coupling as well. • Run the new cable corresponding to the existing cable in the monitor trolley.