Replacing/Installing Software On The Master Option Board; Activating/Deactivating Mob Backup Battery - Siemens SIRECUST 1280 Service Manual

Bedside monitors
Table of Contents

Advertisement

E331 .E2501.061.21.02.02
System SIRECUST 1280/1281
Bedside Monitor - Service Manual
4.9.2.
Replacing/Installing Software
on the Master Option Board
Remove the Master Option Board
following the procedure outlined in
Section 4.9.1.
Remove
the
EPROM
from
J23
using an IC extractor.
Carefully
insert the
replacement
EPROM
into position J23, with the
notche facing toward connectors
X3, X4, X5, and X6 on the left side
of the board.
Firmly
press
on
the
EPROM
to
seat it in the socket.
Reinstall the Master Option Board
following the procedure outlined in
Section 4.9.5.
4.9.3.
Activating/Deactivating MOB
2a.
Backup Battery
Refer to Figure 4-6.
Remove the Master Option Board
following the procedure outlined in
Section 4.9.1.
To
activate
battery
back-up,
remove jumper X7 from between
pins
1
and
2,
and
install
it
between pins 2 and 3.
To
de-activate
battery
back-up,
remove jumper X7 from between
pins
2
and
3,
and
install
it
between pins 1 and 2.
NOTE:
Always return jumper X7 to the
inactive
state
(between
pins
1
and
2)
when the board is not in service.
Install shorting jumper
if 4 Ch. DAC used in LOCATION 2
___
Install shorting jumper
if 4 Ch. DAC. used in LOCATION 1
>
0006
7
V1
V2
Va
Va
WY
LOCATION 1
LOCATION 2
o
EEJX
レグ
СЁ
xa
J12
nl
©
©
"KY
Analog output
Analog outputs
Channels 1-4
Channels 5-8
xo
J23
X4
[ee] x10
7
4
.
2
LOCATION 4
LOCATION 3
J36
J37
7
X12
2
D
©
HHHHH
РА
7
UART signals
EN
ss HAHBESS
J39
Analog Output | X13
x16
%
X2
X1
Install shorting jumper
if 4 Ch. DAC in LOCATION 3
Figure 4-6.
Master Option Board Configuration