Requirements

Replacing damaged or missing screws

As specified in the installation drawings, damaged or missing screws may only be
replaced by steel screws conforming to DIN 267 and having the specified tensile
strength.
CAUTION

Cleaning

The unit must always be switched off or disconnected before cleaning. Use only water to
clean the unit or a lukewarm solution consisting of water and a household cleaning
agent.
Do not use abrasive cleaning agents or organic solvents such as benzine, alcohol or spot
remover.
Do not spray water on the unit.
For additional information, refer to the Operating Instructions "Cleaning/Disinfecting".
Siemens AG
Medical Engineering
All Allen screws must have a tensile strength rating of 8.8.
SPR8-125.034.01