Take snapshots
A snapshot is a frozen frame of preconfigured waveforms or trends
saved in the monitor memory. You can take up to 10 snapshots. It is
automatically numbered. When graphical trend view, snapshots
mark "s" in mark field; when numerical trends view, the number
appears in the column 'Mark'.
For configuration, see
"Configuring snapshots"
the monitor before
use".
To create a snapshot manually:
Take Snapshot
Press the
To create automatic snapshots:
The monitor can automatically take snapshots on Brady, Tachy, Art
sys/dia/mean high, Art sys/dia/mean low, Asystole, V Fib and V Tach
alarms if automatic snapshot creation is enabled. To enable this
function, see
"Configuring snapshots"
monitor before
use".
in section
"Setting up
key.
in section
"Setting up the
Viewing and printing snapshots
To view snapshots:
Pt.Data & Trends
1. Press the
2. Select Trends - Snapshot - Next Snapshot.
Turn the Trim knob to move to the next snapshot. In the waveform
field, you can see the time the snapshot was created. Five fields can
be displayed on the snapshot page, and six fields can be printed.
In the trend field, the graphical trend view or the numerical trends
view are displayed according to the snapshot's configuration.
To print snapshots:
Pt.Data & Trends
1. Press the
2. Select Trends - Snapshot.
3. Select Print Page.
NOTE: Network laser printer only.
NOTE: If the graphical trend time scale is 20 minutes, the displayed
time period is 30 minutes. So after 30 minutes of snapshot creating,
the trends for that time will disappeared.
Erasing trends and snapshots
Admit/Discharge
1. Press the
2. Select Discharge.
If the monitor has been turned off from the ON/OFF key but the
patient has not been discharged, the trend data will be stored in the
memory for 15 minutes.