Temperature
Measurement units
Measurement range
Measurement accuracy
Display resolution
Probe types supported
Temperature self-check
Probe type time response
Reusable skin temperature
probe:
Reusable adult central
temperature probe:
Reusable pediatric central
temperature probe:
Disposable skin
temperature probe:
Disposable central
temperature probe, 12F:
Disposable central
temperature probe, 9F:
° Fahrenheit (F)
° Celsius (C)
10 to 45°C (50 to 113°F)
± 0.1°C without temperature sensor
± 0.1°C at 25 to 45 °C with reusable
probes
Use only GE Healthcare recommend
temperature YSI probes.
At start-up and then every 10 minutes
3 s
6 s
4 s
3 to 6 s
5 to 8 s
5 to 8 s
144
Airway gases
Accuracy specifications apply in normal conditions after a 30 min
warm-up period:
Ambient temperature
Ambient pressure
Ambient humidity
Sampling rate
E-miniC module
E-sCO, E-sCAiO, N-CAiO
module
Warm-up time
E-miniC module
E-sCO, E-sCAiO, N-CAiO
module
Respiration rate
Measurement range
E-miniC module
E-sCO, E-sCAiO, N-CAiO
module
18 to 28°C, within ±5°C of calibration
660 to 1060 hPa, ±67 hPa of
calibration
20 to 80%, within ±20% RH of
calibration
150±25 ml/min (sampling line 2 to
3 m, normal conditions)
120±20 ml/min
1 minute for operation, 30 minutes
for full specification
1 minute for operation with CO
O
and N
O,
2
2
5 minutes for operation of
anesthetic agents,
20 minutes for full specification
4 to 80 breaths/min
4 to 100 breaths/min
,
2