GE B20 User Manual page 150

Patient monitor
Hide thumbs Also See for B20:
Table of Contents

Advertisement

N
O measurement range
2
N
O measurement
2
accuracy
N
O display resolution
2
N
O total system response
2
time
N
O rise time
2
N
O drift
2
Anesthetic agents
Anesthetic agents
specified measurement
range
Anesthetic agents
measurement accuracy
Anesthetic agents display
resolution
Anesthetic agents total
system response time
0 to 100 vol%
0 to 85 vol%: ± (2 vol% + 2% of
reading)
85 to 100 vol%: ± (2 vol% + 8 % of
reading)
1%
< 3.4 seconds with a 2 or 3m sampling
line
< 320 ms with 120 ml/min sampling
rate and 2 or 3m sample line
< 0.3 vol%
Hal, Enf, Iso: 0 to 6 vol%
Sev: 0 to 8 vol%
Des: 0 to 20 vol%
± (0.15 vol% + 5% of reading)
0.01% for values between 0...1%
0.1% for values equal or above 1%
< 3.5 seconds with a 2 or 3m sampling
line (< 3.8 for Hal)
146
Anesthetic agents rise time Hal, Enf, Iso, Des: < 420 ms with a 3 m
Hal drift
Enf drift
Iso drift
Sev drift
Des drift
The E-sCAiO, N-CAiO modules automatically identify the anesthetic
agent present in the sampled gas and measure the concentration
of the identified agent.
Identification threshold
Identification time
The E-sCAiO module automatically identifies mixtures of two
anesthetic agents present in the sampled gas and measures the
concentrations of the two identified agents.
Identification threshold for
the second agent at 1 MAC
of the first agent:
Effects of interfering gases and vapors
E-miniC module
sampling line
Hal: < 800 ms with a 3 m sampling line
Hal, Enf, Iso, Des: < 700 ms with a 6 m
sampling line
Hal: < 1800 ms with a 6 m sampling
line
< 0.1 vol%
< 0.1 vol%
< 0.1 vol%
< 0.1 vol%
< 0.3 vol%
0.15 vol%
< 20 s
0.2 vol% +10% of the concentration of
the first agent