Preparing the patient for a IBP measurement
1. For the setup, prepare the transducer kit according to the
manufacturer's instructions.
2. Remove entrapped air from with the transducer setup.
3. Zero the transducer by opening it to air, pressing the
All
command board key or the
Zero Pressures - Zero ALL. You can zero one channel at a time
Zero IBP1
with selection
NOTE: Selecting Zero ALL does not zero ICP. Zero it separately.
NOTE: The transducer is always leveled to the mid right atrium.
Zero the transducer and pressures whenever the patient's
position is changed.
4. Open the line to the patient.
NOTE: Invasive pressures need to be zeroed after reconnecting the
pressure transducer or cable, and whenever patient's position is
changed. If any channel has not been zeroed, the message 'InvBP's
not zeroed' appears. The invasive pressure alarms advance to
medium and high priority levels regardless of the zeroing.
NOTE: IBP transducers must comply with the requirements of IEC
60601-2-34, must be provided with defibrillator protection, have a
frequency response exceeding 15Hz and contribute not more than
2mmHg to the overall measurement error.
IBP Zero
IBP
monitor key and selecting
Zero IBP2
or
. Zero each channel.
Set up ventilation mode
Respiration causes artifacts in invasive pressures. At the end of
expiration the artifact is at it's smallest.
IBP
1. Press the
key.
2. Select Ventilation Mode.
3. Select Spont for spontaneous respiration; select Contrl for
controlled ventilation.
IBP channel setup
The label of the pressure channel sets its display scale, color, filter,
alarm source and alarm limits. These selections are preconfigured
according to the label you choose. You can adjust these settings
also.
To select the label:
IBP
1. Press the
key.
2. Select IBP1 Setup - Label and IBP2 Setup - Label. For factory
default descriptions, see next page.
97