Network
Warnings
•
Do not use with iCentral software V5.0.3 and earlier.
•
Do not use Mobile Care Server software V5.2 and earlier.
Caution
•
Install HL7 network interfaces as specified, and only by qualified
personnel.
The monitor has been verified to be able to work in CARESCAPE*
Network and S/5 network environments. Other network
infrastructures are not supported.
The monitor has EMR connectivity. There are three ways to acquire
trended vital sign data from patient monitor: HL7 directly from
monitor, HL7 from the CARESCAPE Gateway, or connecting to the
serial port of each monitor. The monitor HL7 (Health Level Seven)
message match with IHE PCD-01 OBR/OBX format.
On the CARESCAPE network, the monitor is able to:
•
communicate with GE CARESCAPE CIC pro version 4.0.8, 4.1.1
and 5.1.0
•
communicate with GE CARESCAPE Central Station v1
•
compatible with Aware Gateway Server v1.6
•
compatible with CARESCAPE Gateway server v1.
•
compatible with Mobile Care Server v6.0
•
compatible with S/5 Collect v4.0
On the CARESCAPE network, the bedside monitor can
simultaneously communicate with a maximum of 4 central stations,
1 Gateway server, 1 Mobile Care Server and up to 1,000 other
monitoring devices . The bedside monitor cannot be the Time
Master on the CARESCAPE network.
On the S/5 network, the monitor is able to:
•
communicate with iCentral version 5.1.1 and 5.1.2.
•
compatible with S/5 Collect v4.0
•
compatible with Mobile Care Server v6.0
The monitor does not support the Patient Data Server (PDS). The
monitor's realtime patient data can't be viewed on other monitors
(e.g. Dash 3000/4000/5000, Solar 8000, B850, B650).
The following monitoring information is transferred to the central
station:
•
Supported parameters
•
Real-time waveforms
•
Real-time alarms
•
Arrhythmia alarms: Asystole, VFib, VTach, Brady and Tachy
•
Graphic and numeric trends
•
Admit/discharge status
•
Patient demographic information
•
Alarm settings