Canon CXDI-1 Operation Manual page 106

Digital radiography system
Table of Contents

Advertisement

8. Calibration
2
Select the sensor unit.
If multiple sensor units can be used, first select a sensor unit by touching that sensor unit
button.
3
Remove the grid unit.
If a sensor unit with removable grid unit is selected in step 2, remove the grid unit.
If [START] is touched while the grid unit is attached, it will display an error, and
calibration will not be performed.
4
Set times (number of exposures).
Set the number using
Or, touch the frame and enter the value from the ten key window.
5
Select the bucky time.
Touch
to select from the list.
This setting can only be made when using a movable grid.
Calibration will not be performed properly unless the X-ray irradiation time is within the
bucky time.
6
Set the exposure parameters.
Set the number using
Or, touch the frame and enter the value from the ten key window or parameter table.
7
Set the timeout for warning.
Set the period from the completion of calibration until a warning is displayed to perform
the next calibration. Touch [hour]/[day]/[week] to select the unit of time.
Touch
or
to set the display duration.
Or, touch the frame and enter the value from the ten key window.
A warning will not be displayed if the time is set to 0.
Warning on calibration
The warning on calibration will be displayed as a yellow [ERR] mark blinking on the
lower right side of the message area.
Although exposures can be performed even when the warning is displayed, you must
remember to calibrate the instrument later, as the warning will disappear when another
error occurs.
8
Enter a memo as required.
Touch the frame to display the software keyboard and enter the memo in 64 or fewer
alphanumeric characters.
or
.
or
.
– 98 –