M3154 Database Server - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

The current workstation for Patient Link is shown in Figure 1-2.
M3154
The M3154 Database Server is a Server that provides database storage of patient monitoring
Database
data for all devices on the Network. Information Center application software provides for
Server
storage of up to 96 hours of patient monitoring data (full disclosure waveforms, physiologic
parameters, alarms, multi-lead arrhythmia, ST segments, and EASI 12-lead presentations)
from 128 patients. Patient monitoring data, including events and trends, can be reviewed by
any M3155 Information Center or M3151 Client on the Network.
The Information Center Web Access option permits viewing of all patient data stored on the
Server (waves, alarms, events, ST segments, trends, and viewable recording strip data) by PCs
on the hospital's HIS intranet. Up to 100 simultaneous users can access stored patient data
using standard web browsers -- Internet Explorer (Release 5.0 or greater) and Netscape
(Release 4.7 or greater).
Standard components for the M3154 Database Server are shown in Figure 1-5 and include
the following:
Options for the M3154 Database Server are:
PC
Figure 1-4 Basic Patient Link Workstation
Server with:
– hard disk drive
– floppy disk drive
– CD ROM and disk drives
– modem (US/Canada only)
– 10BT/100TX Ethernet Network adapter card (for Clinical Network access)
– 10BT/100TX Ethernet Network adapter card (for connection to Hospital LAN)
keyboard
mouse
UPS
Microsoft Windows 2000 Operating System
Information Center application software with database storage of the following data for
up to 128 patients:
– 24 hours of full disclosure waveforms (4 waveforms per patient) and physiologic
parameters
– 150, 30 second alarm records and saved strips, 4 waves per record
– 24 hours of trends with 1 minute of resolution
– 24 hours of ST segments
– 24 hours of events with 1 hour strip function
48, 72, or 96 hours of patient data storage per patient