Adding Network Devices - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

Configuration
Buttons at the bottom of the window do the following:
The remaining buttons to the right of the List of network centrals and hallways have the
same use as for Clinical Units.
List of network devices lists the network devices connected to this system. This field is
accessed by Network Statistics to provide status information on network switches and
access points.
Columns in the List of network devices field are:
Field
Device Name
Host Name
Device Type
IP Address
Printer MAC Address
Note
IP Addresses for devices must be entered in this window in order to identify their
performance in the Network Statistics application.
If no switches are entered, then the Equipment Setup page will not allow any of the Mx/
IntelliVue Patient Monitor installation options.
Adding Network
In addition to adding these network devices to the Database Server/Information Center
Devices
system, the network devices must also be configured to communicate with the system. The
following table provides what their additional configuration requirements are:
Device Type
12-Lead ECG Management
24 Port Switch
6-28
Software Installation & Configuration
OK stores the information entered. It then appears in the List of network devices in
the Network Configuration window.
Cancel closes the Add to network domain window with no Device being added.
Description
Names given to a 12-Lead ECG Management, 24 Port-Switch, Application Server, Holter,
Networked Printer, SNTP Time Source, Alert Data Export Client, Wireless Access Point
(with an identifying icon) on the Network of this Server. Device Names for devices are specified
in the Add new network device to this domain window when the device is added to the
network
Host name of device being added
Type of device: 12-Lead ECG Management, 24 Port-Switch, Alert Data Export Client,
Application Server, Holter, Networked Printer, SNTP Time Source, Wireless Access Point
Internet Protocol Address of the device that identifies its location on the Network.
(Media Access Control Address) is the 12 digit HEX number that uniquely identifies the printer.
This number must be obtained from the printer's configuration printout.
Systems Supported on
Database Server
all
Requirements
ECG Management system must be configured to
receive Information Center data (PMDExport folder
must be created). The Database Server must be
configured with the 12-Lead ECG Management
system's hostname
Must be configured using the Information Center
ConfigTool