M3154 Database Server Installation; Server To Hospital Intranet - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

M3154
The M3154 Database Server should be installed in its intended location. This includes the
Database
Server, display, keyboard, and, mouse. The Server has 4 wheels on its base and is intended
Server
for sitting on the floor. The display, keyboard, and mouse are intended for a work surface.
Installation
Step 1. Install the Database Server components in their intended positions.
Step 2. Install the 1000 VA UPS with the proper voltage for the Net Server as follows:
Power Cord Connector
USB
TVS
Server to
If Information Center Web (Option #C22) has been purchased, interconnect the Database
Hospital
Server to the Hospital's intranet using proper cabling. This is a 100 Mbit/s connection.
Intranet
– Turn Off the UPS On/Off switches.
– Connect the battery wire of the 1000 VA UPS.
– Connect the UPS input power cord to a properly grounded electrical output
– Connect the Server power cord to a UPS battery backup outlet.
– Connect the Philips provided 9 pin UPS communication cable (gray) to the
computer interface port of the UPS and the Serial A port on the rear of the
Server. See Figure 5-15.
Philips 9 Pin UPS/Server
Communication
Cable (gray)
SEN-
SITE
OUTPUT 120-
50/60 Hz
INPU
Figure 5-15 ML370 G3 UPS/Server Communication Cable Connection
Locating System Components
Hospital LAN
Card
(Optional)
Clinical Network LAN
Connector
Parallel Connector
Display Connector
Serial Connector B
Keyboard Connector
Mouse Connector
Hardware Installation
5-15