4. The Adjustment dialog shown in Figure 7.24.1-2 is displayed. Input the calibration value
of horizontal direction (X) and vertical direction (Y) between two points, or the angle
between two points. And then, click the [Next] button.
NOTE : In case of inputting the angle, the value of vertical direction (Y) is calculated and
calibrated by the value of horizontal (X) and input angle.
5. The confirmation message shown in Figure 7.24.1-3 is displayed. When clicking the [OK]
button, [X, Y] calibration is executed. The left side of the [X, Y] button in the [Calibration]
block changes into an orange mark.
X, Y calibration ends with the above-mentioned procedure.
When selecting the [X, Y] from the [Calibration] menu or clicking the [X, Y] button on the
[Calibration] block with X, Y calibration executed, the content of X, Y calibration is
cleared and returned to the default.
Figure 7.24.1-2 Adjustment of X, Y Dialog
Figure 7.24.1-3 Confirmation Dialog