Control Specification - Sanyo MCO-5M Service Manual

Multi-gas incubator
Hide thumbs Also See for MCO-5M:
Table of Contents

Advertisement

Control specification

Model
Heating system
Temperature control
system
Temperature display
CO
controller
2
CO
density display
2
O
controller
2
O
density display
2
Humidifying system
Water level sensor
Air circulation
Alarm
Remote alarm
Self diagnosis function
UV lamp control *
* UV lamp is an optional accessory.
MCO-5M
Direct Heat & Air (DHA) jacket system
PID control
Sensor: Thermister sensor (103AT-1)
Directly detect and control
LED digital display (resolution: 0.1 )
PID control / TC sensor (direct detection in chamber)
LED digital display (resolution: 0.1%)
PID control / zirconia sensor (direct detection in chamber)
LED digital display (resolution: 0.1%)
Natural evaporation with humidifying pan
Optical type
Fan assisted
High/low temperature alarm, CO
Upper limit temperature alarm, water level alarm, UV lamp alarm
When an alarm is triggered:
Remote alarm ON (Alarm interlocks with buzzer)
Allowable contact capacity:DC 30V, 2A
Error code and internal temperature are alternately displayed
Buzzer and remote alarm contact ON
Automatic ON-OFF control (changeable with Function mode)
1. Interlock with door:
UV lamp lights for setting period after door is open and shut.
2. OFF mode: UV lamp does not light.
4
density alarm, Door alarm,
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents