How To Reverse The Door; Uv Lamp (Mco-18Uvs2 *) Installation Procedure - Sanyo MCO-5M Service Manual

Multi-gas incubator
Hide thumbs Also See for MCO-5M:
Table of Contents

Advertisement

How to reverse the door

1. Unplug the unit and ensure the power is not supplied.
Remove the rear panel.
2. Take the inner and outer door off.
(1) Take out power panel which is at the front.(4screws)
(2) Pull out the inner door hinge pin (2pcs), take the
inner door off.
(3) Remove the connecter (2pcs) of the door lead wire,
earth wire and nylon clip. Remove the bushing for
door lead wire outlet and take the lead wire out.
(4) Remove the outer door hinge cap. (4screws)
(5) Remove 8pcs of outer door hinge screw and take
the outer door off.
Do not miss the attachment plate for outer
Note
door hinge.
3. Preparation to reverse the door
(1) Remove the cover of outer door bottom, and pull the lead wire out to the opposite side and
put the cover back on.
(2) Place the outer door hinge and the cap to the opposite side.
Note
Outer door hinge screw fastening torque = more than 80kgf/cm
Cap
Cover for outer
door bottom
Inner door hinge pin
Bushing for door
lead wire outlet
Outer door
hinge cap
Nylon clip
Door lead wire
44
Inner door
hinge pin
Power panel
Details of A
A
Ground
Door lead wire
connecter
Cap