The data to be communication available are follow:
Internal temperature (TEMP. PV)
Internal CO
Internal O
UV lamp turning on period
AT sensor current temperature (AT.PV)
Door open
Humidifying tray water level
Internal temperature SV (TEMP. SV)
Internal CO
Internal O
UV lamp turning on period SV (F01)
F22: Communication parameter set (initial: 000)
<Direction for
Input "F22" on temperature display and press ENT key to show "000" on CO
use>
display. Then change the 3
local mode, and "1" to set in remote mode.
Change the 2
in 9600bps.
Control mode (the 3
Baud rate (the 2
F24: Interlocks the remote alarm with buzzer (initial: 000)
<Direction for
Input "F24" on temperature display and press ENT key to show "000/001" on
use>
CO
display. Then change the 1
2
000: Buzzer/Remote alarm not interlocked
001: Buzzer/Remote alarm interlocked
F25: Ring Back time set (initial: 030 (30min), settable range: 0~60min)
<Direction for
Input "F25" on temperature display and press ENT key to set current Ring
use>
Back delay time. Change the 2
0~60min.
Note) Ring Back time set in F25 = "000" means BZ would not comeback
F26: Displays a coefficient for O
Input "F26" on temperature display and press ENT key to show a coefficient
for O
2
revert to PV display.
F27: Displays a coefficient for O
Input "F27" on temperature display and press ENT key to show a coefficient
for O
2
revert to PV display.
F28: Displays a coefficient for O
Input "F28" on temperature display and press ENT key to show a coefficient
for O
2
revert to PV display.
density (CO
PV)
2
2
density (O
PV)
2
2
SV (CO
SV)
2
2
SV (O
SV)
2
2
rd
nd
digit to "0" to set baud rate in 2400bps, "1" in 4800bps and "2"
rd
digit)
nd
digit)
zero point adjustment (low density) (initial: 000)
2
zero point adjustment. Press ENT key or leave it for 90seconds to
span point adjustment (low density) (initial: 100)
2
span point adjustment. Press ENT key or leave it for 90seconds to
zero point adjustment (high density) (initial: 000)
2
zero point adjustment. Press ENT key or leave it for 90seconds to
21
digit to "0" to set the communication mode in
0: Local (initial)
1: Remote
0: 2400bps
1: 4800bps
2: 9600bps
st
digit to "000" to set the interlock mode.
nd
digit to set delay time in the range of
2

Hide quick links: