Sanyo MCO-5M Service Manual page 55

Multi-gas incubator
Hide thumbs Also See for MCO-5M:
Table of Contents

Advertisement

Setting procedure of Control panel.
When use MCO-5GC(the automatic switcher of CO
MCO-5M automatic gas switching mode according to the following procedures.
Description of operation
1
Turn the power switch ON.
2
Press CAL key for 5 seconds.
3
By pressing
set the figure to F03.
4
Press ENT key.
5
By pressing
set the figure to 384
6
Press ENT key.
7
Press CAL key for 5 seconds.
8
By pressing
set the figure to F08
9
Press ENT key.
10
By pressing
set only the left digit figure to 1.
NOTE:Don't
digit figure and right
digit figure.
11
Press ENT key.
operated
key and
key,
key and
key,
key and
key,
key and
key,
change
center
gas supply line),it is necessary to set the
2
Key
Indication after operation
The current chamber temperature is
-
displayed.
The left digit in the temperature indicator
CAL
is flashed.
In CO
density indicator,
2
ENT
"000" is displayed and the
right digit is flashed.
The current chamber temperature is
ENT
displayed.
The left digit in the temperature indicator
CAL
is flashed.
In CO
density indicator,
2
ENT
"0**" is displayed and the
right digit is flashed.
Setting of MCO-5GC is finished.
ENT
The current chamber temperature is
displayed.
CO
gas supply line indicator A is lighted.
2
52
F 0 3
F 0 3
0 0 0
F 0 3
3 8 4
F 0 8
F 0 8
0 * *
F 0 8
1 * *
"*" means "0" or "1".