Sanyo MCO-5M Service Manual page 50

Multi-gas incubator
Hide thumbs Also See for MCO-5M:
Table of Contents

Advertisement

6. Replace the duct. Place the humidifying tray,
then cover the tray with the humidifying tray
cover as shown in Fig. 5
Humidifying tray cover
7. Replace the insulation into the UV lamp lead wire
hole with the wire passed through the cut on the
insulation. See Fig.6
Fig.6
Fig.5
47
8. Remove a cap shown in Fig. 7.
By using 4 screws enclosed with this kit,
fix the ballast/glow ass'y as shown in Fig. 7.
④Glow ass'y
⑤Ballast
③Screw
Cap
Fig.7