MCO-18UVS2 Installation Procedure
Kit for MCO-18UVS2
1 UV lamp
1 UV lamp cover
6 screws
Procedure
1. Unplug the unit and make sure that power is
not supplied to the unit.
2. Remove the duct from inside the unit, then
take off 2 caps and 2 screws shown in Fig. 1.
Cap
3. Take off the rear cover shown in Fig. 2.
Fig.2
1 ballast & glow assy
4
Screw
Fig.1
4. Take out two insulations for the UV lamp
lead wire hole shown in Fig. 3.
The insulation can be taken out easily
by pushing it from inside of the
chamber. Keep the insulation for the
future use again.
5. Set the UV lamp to the hole and make
sure to be fitted properly.
Fix the UV lamp cover with 2 screws
enclosed in this kit.
Screw
46
1 UV label
1 UV caution label
1 UV door switch
caution label
2 Name plate
1 Patent label
Fig.3
UV lamp
UV lamp cover
Fig.4