Control specification

Model
Heating method
Temperature control
system
Temperature indication
CO
measuring system
2
CO
level indication
2
Humidifying system
Inside air circulation
Alarm system
Door switch
Remote alarm
Self diagnosis function
UV lamp control *
* UV lamp is optionally provided.
MCO-18AIC
Direct Heat + Air jacket (DHA)
Microprocessor PID
Sensor: temperature sensor (103AT-1)
Detect and control the inside temperature directly.
Digital display (resolution: 0.1 C)
Infrared CO
sensor, PID control
2
Digital display (resolution: 0.1%)
Natural vaporization by water in humidify pan.
Humidify pan: W293 x D407 x H41mm
Breeze circulation
High/low temperature alarm, high/low CO
Both display for temperature and CO
intermittently after 15 min.
Independent alarm for overheat
Door lamp
When outer door is opened:
Inside fan OFF, CO
2
(only when door is kept opening for 1 min. or more)
When an alarm is triggered:
Remote alarm ON (Alarm and buzzer are interlocked.)
Contact output: rating of contact 30VDC, 2A
Error code and internal temperature are displayed alternately.
Buzzer and remote alarm contact ON
Automatic ON-OFF control (changeable with Function mode)
1. Interlocked with the door:
UV lamp ON for 5 min. after every door closed.
2. OFF mode (UV lamp does not turn on)
4
level alarm
2
are flashing, and buzzer sounds
2
valve OFF, (UV lamp * OFF) and heater OFF