Technical Alarms (INOPs)
Alarm Text
MORE BED ALARMS
Source - Patient Monitor
NO ARRHYTHMIA
This condition
Note—
displays as All ARRH
ALRMS OFF at the IIC.
Source - Information
Center
NO ECG AT CENTRAL
Source - Patient Monitor
NO ECG SOURCE
Source - Telemetry
5-14
Alarms
Draft - 1 Aug 08
Technical Alarms (Inoperative Conditions)
Priority
Condition
Red or
The monitor is associated
Yellow or
with a transceiver and is
Hard
sending data to the
Technical
Information Center via the
Alarm
transceiver. There are
currently more alarms at
the bedside than can be
transmitted to the
Information Center.
Soft
All basic ST/AR ECG
alarms have been turned
off.
Hard
The ECG measured with
the monitor is not being
sent to the Information
Center via the transceiver.
Hard
The transceiver is paired
with a monitor but the
Information Center is not
detecting a valid ECG
signal from either of them.
What to do
For more information only.
Appears at Information
Center only.
Turn arrhythmia alarms on.
Check equipment
connections and
configuration settings.
Check equipment
connections and
configuration settings.