Verifying Electrode Connections
6-26
ECG Monitoring
Draft - 1 Aug 08
Step
Action
2
Expected Response:
If 3-wire cable is attached: Red, White and Black indicators
illuminate, then all turn off.
If 5-wire cable in Standard mode is attached: Red, White,
Black, Green & Brown indicators illuminate, then all turn off.
If 5-wire cable in EASI mode is attached: Red, White, Black,
Green & Brown indicators illuminate, then all turn off. The
EASI indicator also illuminates briefly.
If 6-wire cable is attached: Red, White, Black, Green, Solid
Brown and Brown/White indicators illuminate, then all turn
off.
If all leads are off, the green indicator will be lit.
3
Unexpected Response:
Any other response indicates a problem with the transceiver or lead
set.
Check the leadset connection. The colored line should not be
visible; otherwise the wrong leadset type may be detected.
If the INOP, INVALID LEADSET, and the corresponding
audible tone occur, the leadset inserted does not match the
transceiver's configuration. Insert the correct leadset per the
configuration.
Use a new leadset.
If the problem is not corrected, contact your service provider.