Telemetry Functions at the Information Center
Unit Settings - Telemetry Setup
Control
Function
SpO
Alarm
Turn SpO
2
Information Center.
SpO
Limits
Increment/decrement SpO
2
High
alarm limit by 1 (in %).
SpO
Limits
Increment/decrement SpO
2
Low
alarm limit by 1 (in %).
Unit Settings - Default Leads
Control
Function
Leadset
Set the unit default leadset type.
Allowed
3-wire
Set the unit default lead.
5-wire, ECG1
Set the unit default lead.
5-wire, ECG2
Set the unit default lead.
9-14
Telemetry Functions at the Information Center & TeleMon
Draft - 1 Aug 08
alarms on/off at the
2
high
2
low
2
Settings
enable (on)
disable (off)
Limit maximum is 100. Limit
minimum is 51 (adult) or 31
(pediatric). High and low
limit must be at least 1%
apart.
Limit maximum is 99. Limit
minimum is 50 (adult) or 30
(pediatric). High and low
limit must be at least 1%
apart.
Settings
All, EASI, No EASI, 3 wire
I, II, III
I, II, III, MCL, AVR, AVL,
AVF, V
I, II, III, MCL, AVR, AVL,
AVF, V
Factory
Default
enable
100 (adult,
pediatric)
90 (adult,
pediatric)
Factory
Default
All
II
II
V