Transceiver Operation when Connected to TeleMon

When connected to TeleMon, the following differences in transceiver operation
will be in effect.
Important—
connected to TeleMon.
Alarms
Note—
TeleMon silences the alarm at TeleMon only. It has no effect on the alarms at
the Information Center.
ECG Operation
Draft - 1 Aug 08
The transceiver must have good batteries installed before being
Alarm Suspend/Pause and Standby can be initiated only from TeleMon,
not from the transceiver or the Information Center. The alarm suspend
indicator on the transceiver will accurately reflect the current state of
alarm pause.
If the transceiver is connected to TeleMon, silencing an active alarm at
Vb, the second V-lead in 6-wire lead-set, is not supported.
After a change in leadset, the TeleMon returns to the default ECG
settings, and arrhythmia relearn occurs automatically. Be sure to check the
monitoring leads after you switch leadsets. See "Relearning" on page 6-
28.
Telemetry Functions at the Information Center & TeleMon

Transceiver Operation when Connected to TeleMon

9-19