Unit Settings - NBP Setup
Control
Function
Systolic Low
Increment/decrement NBP low
alarm limit by 1.
Diastolic
Increment/decrement NBP high
High
alarm limit by 1.
Diastolic Low
Increment/decrement NBP high
alarm limit by 1.
Mean High
Increment/decrement NBP high
alarm limit by 1.
Mean Low
Increment/decrement NBP high
alarm limit by 1.
Draft - 1 Aug 08
Settings
Limit Minimum is 28 (Adult)
Limit Maximum is 255
Limit Minimum is 28 (Ped.)
Limit Minimum is 90 (Adult)
Limit Maximum is 260
Limit Minimum is 45 (Ped.)
Limit Minimum is 28 (Adult)
Limit Maximum is 40
Limit Minimum is 28 (Ped.)
Limit Minimum is 65 (Adult)
Limit Maximum is 260
Limit Minimum is 55 (Ped.)
Limit Minimum is 28 (Adult)
Limit Maximum is 50
Limit Minimum is 28 (Ped.)
Telemetry Functions at the Information Center & TeleMon
Telemetry Functions at the Information Center
Limit Maximum is 155
Limit Maximum is 260
Limit Maximum is 85
Limit Maximum is 260
Limit Maximum is 60
Factory
Default
90 Adult
70 Pediatric
90 Adult
70 Pediatric
50 Adult
40 Pediatric
110 Adult
90 Pediatric
60 Adult
50 Pediatric