The connection with CXDI-50G/50C and X-ray generator equipment
Connection conditions
1. The X-ray exposure signal line must be insulated, and its total impedance must be 100 ohms or
less.
2. The maximum contact voltage of the X-ray exposure authorization signal line is AC 250V and
DC 30V, and its current ranges from 10mA to 2A.
Only the secondary power supply can be connected.