Canon CXDI-50C Service Manual page 147

Cxdi-1 system
Table of Contents

Advertisement

(5) Attaching cable clamp plate
Place the cable clamp plate on the cable, and lightly tighten the sensor side screw first,
and then the outer side screw. Make sure that the cable clamp plate is horizontal.
Sensor side screw
Sensor side screw
(6) Tightening cable clamp plate screws 1
Tighten the sensor side screw until the space between the screw and the cover is 2 mm. At
this time, cable clamp plate will be inclined as shown below.
2mm
‚ Q ‚ ‚
Technical Information
Space between the sensor side cable clamp plate and cover is made smaller first in order to
prevent the cable on the sensor side being caught in the cable clamp plate.
4. Repair Guide
12
Outer side screw