(3)
Cable connections
1) Loosen the lock nut for each cable and connect the cables to the power box. Then fix
the cables by tightening the lock nuts.
Note: To avoid the risk of damage when the cables are removed with very large force, Check
tightening torque of the bush (refer the following figure).
2) Attach the cable clamp to the sensor cable, and then fix it to the power box using the
screw (M4 x 6 mm).
Screw (M4×6mm)
2. Installation Manual
Sensor cable
X-ray I/F cable
Gap 1.0 to 1.5mm
<Sensor cable>
- 12 -
Lock nut
Remote cable
No gap
<X-RAY I/F cable,
Remote cable>
Clamp the cable's
shield portion
Cable clamp