Exchanging the sensor cable
2.4
(1) Removing the Cable
Screw M2.6x6 (x2)
Cable clamp 2
Bend the power line.
Disconnect the power
cable connector.
Clamp cushion
1. Remove the screws (M3x8) x2 that fixes the cable clamp 1.
2. Remove the screws (M2.6x6) x2 that fixes the cable holder.
3. Remove the screws (M2.6x6) x2 that fixes the cable clamp 2. Do not lose the screw
because it is held by the clamp cushion.
4. Disconnect the connectors from PCB-50LED and PWB-50Di.
5. Remove the sensor cable unit.
NOTES:
1. During assembly, lay out the signal cables carefully to they will not get caught by
covers.
4. Repair Guide
Screw M3x8 (x2)
Screw M2.6x6 (x2)
7
Cable clamp 1
Bush
Cable holder
Be careful not to get the cable
caught between covers.
Disconnect the signal
cable connector.
The cable protect sheet is put
under the cable so as to protect
the sensor cable.
Sensor cable