Figure 5-1 Installation Diagram-Cat. 4909 Cells With Junction Box Head Connected To 7082; Conductivity/Resistivity Analyzer - Honeywell 4909 CPVC Operation Manual

Conductivity cell removal/insertion assembly
Table of Contents

Advertisement

4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual
RED
B
Temp.
Comp.
D
GREEN
C
WHITE
Cell
BLACK
A
Internal Cell Assembly
Configuration
NOTES:
1. For pure water samples in non-conductive (plastic, glass, etc.) piping, ground the black cell electrode lead near the cell. Alternatively,
connect to the 7082 ground screw as shown dotted. Do not ground 10, 25, or 50 constant cells.
2. 70 8 2- 1 6, 17 , 1 8 , 1 9 (only)
Use 22 gage minimum coaxial cable type RG59/U connecting shield to terminal "SH" only.
3. 70 8 2 -1 6 , 1 7 , 1 8 , 1 9
For cable runs of up to 500 ft., use: 18 gage minimum, three conductor cable.
For cbale runs of 500 - 1000 ft., use: 16 gage minimum, three conductor cable.
7 08 2 -1 3 , 14 , 1 5 [coax and shield (SH) not used]
For cable runs of up to 500 ft., use: 18 gage minimum, four conductor cable.
For cable runs of 500 - 1000 ft., use: 16 gage minimum, four conductor cable.
4. Cell to analyzer cables are considered low level. Run seperate from high level wiring.
Figure 5-1 Installation Diagram-Cat. 4909 Cells with Junction Box Head
20
4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual
Four Point Terminal Plate
with #6-32 Screw Terminals
D
A
B
C
View of Junction Box Head
with Cap Removed
Ce l l A s s e mb l y Co n n e ct i o n s
I
II
III
IV
4905
-
-
333 - X1
IV
I
II
III
4973
4974
-
333 - X1 -
I
II
III
IV
4908
-
-
333 - X1
4909
Connected to 7082 Conductivity/Resistivity Analyzer
1000 ft. max.
Note 3
B
D
Coax Cable Shield
A
C
Note 2
Note 1
V
VI
An a l y ze r I np u t C on n e c ti o n s
V
VII
V
VI
Five Point Terminal Board.
Each Terminal Will Accept
#16 Gage Max. Wire
W K SH G
R
Conductivity/Resistivity
Analyzer
a/n 23345
GND
7/99