Replacement Parts; Table 4-1 Replacement Parts; Table 4-2 Voltage And Time Limits For Platinizing Cells - Honeywell 4909 CPVC Operation Manual

Conductivity cell removal/insertion assembly
Table of Contents

Advertisement

4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual

4.3 Replacement Parts

Complete Assembly
Cell Assembly
Complete CPVC Removal Device
Universal Head Parts
Platinizing Solution (3 oz. bottle)
Cell Guard Tube (0.01, 0.1, and 1.0 Constants Only)
Support Tube, 1/2" NPT Sch. 80, CPVC (12" Immersion)
Support Tube, 1/2" NPT Sch. 80, CPVC (6" Immersion)
Junction Box
Cell Extension Leadwire (Table III other than 333)
Three conductor 18 gage cable PVC (105 C max.)
Three conductor 18 gage cable Tefzel (150 C max.)
9782/7082 Standard Ranges
Up to 500 ft.
Three conductor, 18 gage cable (Belden 9493)
and Coax Cable (Belden 9259)
Up to 1000 ft
Four conductor (3 used), 16 gage cable,
Belden 9494 or equivalent and Coax Cable (Belden 9259)
9782/7082 Wide Ranges
Up to 500 ft - Four Conductor, 18 gage
Up to 1000 ft - Four Conductor, 16 gage
DC Voltage
18
4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual

Table 4-1 Replacement Parts

Description

Table 4-2 Voltage and Time Limits for Platinizing Cells

Cell Constant
5
1.5
----
3.0
200 sec
240 sec
6.0
80 sec
100 sec
12.0
----
10
20
25
----
----
----
----
----
180 sec
200 sec
----
120 sec
150 sec
Part Number
See Figs. 1-5 and 1-6
Cat. 4908
31074357
See Fig. 1-6
31103011
065602
074344
074343
31316260
834059
834086
834059
835024
834055
835024
31834052
834055
50
----
----
300 sec
240 sec
7/99