Press C Yes to zero the CO
Or
Press A No to bypass the CO
step #5. If the detector is passcode protected, refer to
code Protect Activated
step #4.
If C Yes is pressed, the following screen displays.
a Warning
Use nitrogen (N
) to zero the CO
2
Apply N
to zero the CO
sensor. The following screen
2
2
displays.
sensor.
2
zero and proceed to
Auto Span
2
Pass-
sensor.
2
Calibration and Setting Alarm Setpoints
flashes whiles the detector zeros the CO
sensor (approximately 30 seconds).

Passcode Protect Activated

4. When auto zero is complete and if the passcode protect option
is enabled, the following screen displays.
The passcode must be entered to proceed.
Press G or H to scroll to the correct passcode and then press
C within 5 seconds to confirm. If entered correctly, the detec-
tor beeps twice and proceeds to
Incorrect Passcode: If the passcode is incorrect or is not confirmed
within 5 seconds by pressing C, the following screens display.
The detector saves the calibration and returns to normal operation.
GasAlertMicro 5/PID/IR
2
Auto Span
step #5.