Applying Gas to the Sensors

The calibration cap, single gas calibration cap, and hose are shipped
with the detector. Refer to
Figure 3.
The calibration cap and single gas calibration cap must only be
used during the calibration span process.
Figure 3. Applying Gas to the Sensors
46
and
Table 11.
for installation.
Note
Table 11. Applying Gas to the Sensors
Item
1
Detector with calibration cap
2
Calibration hose
3
Regulator and gas cylinder
4
Toxic 2 sensor position
5

Single gas calibration cap

Single Gas Calibration Cap
If an O
or ClO
sensor is located in the Toxic 2 position
3
2
(refer to
Figure 3.
and
cap must be used to ensure accurate calibration.
To calibrate O
and ClO
sensors using the single gas calibration cap,
3
2
refer to
Figure
4.,
Table
12., and complete the following procedures.
1. Insert the cap into the Toxic 2 sensor position on the detector
(refer to
Figure
3.). Press firmly until the release tabs click.
2. Connect the calibration hose to the gas cylinder and to the
intake inlet on the cap.
3. Proceed to
Calibration
Description
a Caution
Table
11.), a single gas calibration
Procedure.