Start Calibration; Auto Zero And Oxygen (O 2 ) Sensor Calibration; Zero Co 2 (Gasalertmicro 5 Ir Only) - Honeywell GasAlertMicro 5 User Manual

Gas detector
Table of Contents

Advertisement

GasAlertMicro 5/PID/IR
User Manual

Start Calibration

Verify that the calibration gas being used matches the span
concentration value(s) that are set for the detector. Refer to
Span Gas
Value. Correction factors are not applied during cali-
bration. Correction factors that were set prior to calibration are
restored when the detector returns to normal operation.
Calibrate only in a safe area that is free of hazardous gas
in an atmosphere of 20.9% oxygen.
1. To enter calibration, press and hold C and H simultane-
ously as the detector beeps, flashes, and vibrates to the
corresponding countdown.
Auto Zero and Oxygen (O
2.
flashes while the detector automatically zeroes the
toxic and combustible sensors, and calibrates the O
Depending upon the detector being calibrated, the auto zero
screens display differently.
48
Note
a Caution
) Sensor Calibration
2
GasAlertMicro 5/PID
Do not apply calibration gas during this process, otherwise the
auto zero step will fail.
Zero CO
(GasAlertMicro 5 IR only)
2
If calibrating using the GasAlertMicro 5 or the GasAlertMicro PID, pro-
ceed to
Auto Span
ceed to
Passcode Protect
3. For GasAlertMicro 5 IR models, there is a second auto zero
that is performed for the CO
When auto zero for the other sensors is complete, the following
screen displays.
sensor.
2
GasAlertMicro 5 IR
Note
step #5. If the detector is passcode protected, pro-
step #4.
sensor.
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gasalertmicro 5pidGasalertmicro 5ir

Table of Contents