In the following order, press and continue to hold each button until
Tech mode displays below the Language option.
1. Press and hold H with the right index finger.
2. Press and hold G with the right middle finger.
3. Press and hold C with the left thumb.
4. Press C to enter Tech mode. The options are as
follows:

Sensors

Initialize
Force Calibration
Bump Daily
(Bump test daily)
Stealth Mode
Sleep Mode
(Force calibration)
Sensors
To reconfigure the sensor type, physically change the
sensor prior to entering Tech mode.
When a toxic sensor is physically removed and replaced by another toxic
sensor, the detector must be reconfigured to recognize the change.
If a sensor is replaced, the detector will classify the sensor as
overdue for calibration. Calibrate the new sensor immediately.
Press H to scroll to Sensors. Press
1.
C within 20 seconds to confirm and
access the toxic sensor menu.
Press G or H to scroll to Toxic 1
2.
or Toxic 2 and press C within
20 seconds to confirm.