Honeywell Envitec AlcoQuant 6020 plus Operating Instructions Manual page 33

Table of Contents

Advertisement

Phase
Display on
device
Measurement Warning
Memory full
Measurement
Connecting
/ Printer Test
Printer
Measurement Error
Printing
Printer Test
Error
Wireless
Module
Printer Test
No
Connection
Doc no.: 059-07-1001779_A6020plus_WL_GB-5.2.1 / 12.2021
Result
LED
signals
Confirm
YELLOW
with key
Printout
YELLOW
does not
Flashing
start
Printout
YELLOW
does not
start
Printout
RED
does not
start
Printout
YELLOW
does not
start
AlcoQuant 6020 plus
Cause/solution
The measurement memory is
100% full. The oldest
measurements are overwritten
from now on (ring memory).
Download data with PC software
and/or delete old measurements.
(Option Wireless)
The wireless module of AlcoQuant
6020 plus tries to establish a
connection to the printer.
If the connection attempt takes
more than a few seconds, check
>Printer Setup< to confirm that the
Connection ID corresponds with
the Connection ID of the used
printer.
While printing, move device closer
to the printer.
(Option Wireless)
The wireless module of the
AlcoQuant 6020 plus can not
establish or has lost the
connection to the printer.
Check >Printer Setup< to confirm
that the Connection ID
corresponds with the Connection
ID of the used printer.
(Option Wireless)
The built in wireless module of
AlcoQuant 6020 plus is not
accessible.
Restart device. If error persists,
contact service.
(Option Wireless)
The wireless module of the
AlcoQuant 6020 plus cannot
establish or has lost the
connection to the printer.
Check >Printer Setup< to confirm
that the Connection ID
corresponds with the Connection
ID of the used printer.
33