Press the ON/OFF key to return to active mode.
The values measured in the passive mode do not correspond to the breath alcohol
concentration in the deep lung air and therefore do not lead to any conclusions on
the blood alcohol concentration. The exact breath alcohol concentration in the lung
air can be determined exactly only in the active device mode.
Passive-auto-trigger mode (P-AT) – limited use
Depending on the customer specific configuration, the device could be equipped with
the optional measurement mode >>passive auto trigger<<. In this mode the passive
mode sample taking can be triggered by a short blow in, instead of pressing a button
to start the sample. Since this this mode is taking only a quick breath sample instead
of deep lung air with long blow in time, this mode is much faster for mass control
measurements, but not as accurate as an >>Active mode<< test.
It's strongly recommended to use a mouth piece for this mode to improve the
accuracy. If the alcohol limit is zero alcohol this mode must not be used, since very
small breath alcohol concentrations need to be tested at deep lung air to be
detected.
24
The yellow LED is flashing during the evaluation of the
breath
sample.
concentration this can take several seconds.
No alcohol detected
If no alcohol is detected, a high signal sounds off and
the green LED lights up.
Alcohol detected
If alcohol is detected, the device beeps low twice and
the red LED illuminates.
Doc no.: 059-07-1001779_A6020plus_GB-5.2.1 / 12.2021