Offline Configuration - Honeywell SmartLine SLN700 User Manual

Level transmitter non-contact radar hart option
Hide thumbs Also See for SmartLine SLN700:
Table of Contents

Advertisement

Offline Configuration

Overview
Offline Configuration refers to configuring a device when the device is not physically present or
communicating with the application. This process enables the user to create and save a
configuration for a device, even when the device is not there physically. Later when the device
becomes available with live communication, the same configuration can be downloaded to the
device. This feature enables the user to save on device commissioning time and even helps the
user to replicate the configuration in multiplicity of devices with lesser efforts.
The following are the tasks that the user needs to perform for importing offline configuration in
FDM application software and then downloading it to the device.
Create offline configuration template in FDM
Save the configuration in FDM in .fdm format.
Download the offline configuration to the device
For details on creating and using offline configuration, refer to section
Note:
Offline configuration in FDM User's Guide.
Refer FDM User guide for more details. Below is FDM user guide link
https://www.honeywellprocess.com/library/support/Public/Documents/User%27s-
Guide-EP-FDM-11511C.pdf
Importing offline configuration
Using this feature the user can import offline configuration template. The offline configuration
template has to be created in FDM and saved in FDM format.
In case if the offline configuration template is already imported, an
Note:
overwrite message appears.
Revision 1