Analog Output Calibration; About This Section; Equipment Required; Analog Output Signal Calibration Trim - Honeywell SmartLine SLN700 User Manual

Level transmitter non-contact radar hart option
Hide thumbs Also See for SmartLine SLN700:
Table of Contents

Advertisement

4 Analog Output Calibration

About This Section

This section provides information about how to calibrate a transmitter's analog output circuit.
All procedures in this manual assume the transmitter is configured for Loop Current
Mode enabled, see section 4.3.2.

Equipment Required

Digital Voltmeter or milli-meter with 0.02 % accuracy or better
250 ohm resistor with 0.01 % tolerance or better

Analog Output Signal Calibration Trim

With a transmitter in its constant current source mode, its analog output circuit can be trimmed at
its 0 (zero) % and 100 % levels. It is not necessary to remove the transmitter from service.
The user can calculate milliamperes of current from a voltage measurement as follows:
Dc milliamps = 1000 X voltage/resistance
IMPORTANT: Be sure that the accuracy of the resistor is 0.01% or better for current
measurements made by voltage drop