12.2.3. Power Requirements
Voltage:
AC 100V/110V/120V|200V|220V/230V/240V
Frequency:
50Hz/60Hz
Rated Momentary Line Current (shorter than 5s):
•
30A (100V/110V/120V)
•
28A (200V)
•
25A (220V/230V/240V)
Rated Continuous Line Current:
•
20A (100V/110V/120V)
•
12A (200V)
•
10A (220V/230V/240V)
Maximum line impedance: 0.3
Max Line Regulation (Percent): 15% at maximum radiographic exposure.
Line regulation is based on measurements made at the input (primary winding) of the isolation transformer
under standby conditions and at maximum radiographic exposure by the following equation:
Line regulation = 100 (Vn - V1)/V1
Where Vn = no load voltage (standby), and V1 = loaded voltage (maximum radiographic exposure).
Technique factors that constitute the maximum line current condition is 110kV, 20mA.
16A power supply is required in China. And 20 A is required in Japan.
NOTE
|0.6
|0.6