5 Maintenance
5.3 Cleaning procedures

5.3.2 Clean the UV flow cell

5.3.2
Clean the UV flow cell
Maintenance interval
Clean the UV flow cell every six months, or when required.
Required material
The following materials are required:
Procedure
Follow the steps below to clean the UV flow cell.
Step
1
98
NOTICE
Keep UV flow cell clean. Do not allow solutions containing dissolved
salts, proteins or other solid solutes to dry out in the flow cell. Do
not allow particles to enter the flow cell, as damage to the flow cell
may occur.
Luer-to-SNAP connector (see
Waste container
Syringe, 25 to 30 mL
Detergent solution (e.g., Decon™ 90, Deconex™ 11, or RBS™ 25), diluted according
to the manufacturer's instructions.
Distilled water
Action
Disconnect the tubing from the bottom of the UV monitor and replace it with
a Luer-to- SNAP connector.
Luer-SNAP 3.2 connectors, on page
ÄKTA pilot 600 User Manual 29274559 AA
25)