Reset The Pressure Sensors - GE AKTA pilot 600 User Manual

Table of Contents

Advertisement

5.4.1

Reset the pressure sensors

Introduction
The ÄKTA pilot 600 system is fitted with three pressure sensors, for monitoring system
pressure, pre-column pressure and post-column pressure.
The pressure sensors should be reset at intervals by setting the reading for atmospheric
pressure to zero in UNICORN software. The pressure sensors themselves are not acces-
sible to the user.
This section gives instructions for resetting the pressure sensors.
Maintenance interval
The pressure sensors should be tested every week, or when the the ambient temperature
has changed by more than 5°C. Reset the pressure sensors if the atmospheric pressure
reading is outside the range ± 0.02 MPa (0.2 bar, 2.9 psi).
Tubing connections
Resetting pressure sensors involves disconnecting tubing from the appropriate module
according to the table below.
Note:
Monitored pres-
sure
System pressure
ÄKTA pilot 600 User Manual 29274559 AA
Take care to avoid spillage when disconnecting tubing.
Placing
At the outlet from the flow
restrictor
5.4.1 Reset the pressure sensors
Disconnect tubing
Flow restrictor outlet
5 Maintenance
5.4 Calibration procedures
101