6 Performance tests

6.1 When to run performance tests

6.1
When to run performance tests
Schedule
Performance tests may be run at any time, as required. Recommendations are given in
the following table.
Occasion
After system installation
After module addition or replace-
ment
After valve membrane replace-
ment
When a problem is suspected
1
120
Note:
Valve modules are tested for leakage only, not for valve switching function.
Test
Run the system tests in the following order
1 System leakage test
2 System backpressure test
3 System test
If any of the tests fails, correct the cause of the
failure before continuing.
Run the test(s) associated with the specific module.
Run the test for the specific valve.
Test the modules that may be involved in the
1
problem.