2D Mode (B Mode) Checks - GE Vivid 7 Service Manual

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FC091194, R
IRECTION
4-3-4

2D Mode (B mode) Checks

4-3-4-1
Introduction
The 2D Mode is the system's default mode.
4-3-4-2
Preparations
Use a phantom when doing these tests.
1.) Connect one of the probes, to the scanner's left-most active probe connector.
-
-
2.) Turn ON the scanner.
-
The 2D Mode window is displayed (default mode).
3.) If needed, adjust the Display's Brightness and Contrast setting
page
02
EVISION
Figure 4-11 2D Mode Screen Example
See
"Probe Connection" on page 3-13
See
Chapter 9 -
, section
See
4-2-1 - Power ON/ Boot Up
6-2.).
Chapter 4 - Functional Checks
for info about connecting the probes.
9-17
for available probes.
on page
4-2
for info about booting the scanner.
V
7 / V
7 PRO S
IVID
IVID
ERVICE
Status Window
Control Menu
(See "Monitor Adjustments" on
M
ANUAL
4 - 17