Image Quality Tests

4-3-3-1

2D Mode (B Mode) Checks

4-3-3-1-1
Introduction
The 2D Mode is the system's default mode.
4-3-3-1-2
Preparations - 3SC-RS Probe
1) Turn ON the Venue Fit™.
2) Connect 3SC-RS probe
4-3-3-1-3
Flashlight Noise Test (3SC-RS Probe)
3) Select Cardiac preset and run scanning on air
4.) Scan in B mode, increase depth to max, increase gain till image is visible
5) Verify image is homogeneous and no flashlight noise appear
6.) Decrease depth step by step till minimum
4-10
Figure 4-44 2D B-Mode Image Test